Circle Absorber
Construction
The essential
features of the circle absorber are: a carbon dioxide absorber canister
(C), breathing bag (B), unidirectional inspiratory (Vi) and expiratory
(Ve) valves, fresh gas supply (F) and pressure-relief valve (V).
The absorber is connected to the patient via corrugated hoses and
a Y-piece (not shown) attached to the inspiratory and expiratory
valves (Vi and Ve).
The position of the breathing bag and pressure-relief
valve may vary in relation to the absorber, but the above is a common
and satisfactory arrangement.
Function
Inspiration:
Inspiration causes the expiratory valve to close and gas flows from
the breathing bag to the patient via the inspiratory limb of the
circuit. Anesthetic is taken up from the in-circuit vaporizer (VIC),
if fitted.
Expiration: The inspiratory valve closes and gas flows into
the breathing bag via the expiratory limb. Carbon dioxide is absorbed
by the soda lime canister. Excess gas is vented when necessary via
the pressure-relief valve.
Closed and
semi-closed
The circle absorber
may be used as a closed or semi-closed system:
- Closed
systems: the pressure relief valve is closed so that no gas
escapes from the system. Oxygen flows into the system to replace
that consumed by the patient and exhaled carbon dioxide is absorbed
by the soda lime.
The advantage of closed systems is that anesthetic and oxygen
consumption, and atmopsheric pollution, are minimized.
The disadvantages are:
(a) the system is inherently unstable, in that if the fresh gas
flow is not matched exactly to the patient's oxygen consumption,
the system will over-fill or empty, and the patient will be unable
to breathe.
(b) the fresh gas flow rate is usually too small to allow use
of a precision, out of circuit vaporizer.
- Semi-closed
systems: the pressure relief valve is opened allowing excess
gas to escape from the system. This allows higher fresh gas flow
rates to be used.
The
advantages of semi-closed systems are:
(a) The system is more stable in that, if the system fills to
capacity, the excess gas is simply lost via the pressure relief
valve.
(b) The higher flow rates allow use of a precision, out of circuit
vaporizer.
The disavantage is increased anesthetic and
oxygen consumption and atmospheric pollution.
Canister
gas flow
Most modern
canisters are orientated vertically, which facilitates changing
the absorbant and avoids the problems of channelling encountered
with horizontal canisters (as described under the Waters'
Canister). In most absorbers, gas flows from the top to the
bottom of the canister, although in some gas flows from the bottom
to the top.
The
advantage of top-to-bottom flow is that flow tends to be preferentially
directed down the walls of the canister (owing to the lower resistance
of absorbent granules located against the smooth wall of the canister),
so color-change due to exhaustion of the absorbant is easily detected.
Conversely, the color change in bottom-to-top flow canisters starts
in the middle of the absorber where it cannot be seen.
Furthermore,
in bottom-to-top flow absorbers, condensed water vapor tends to
accumulate at the bottom of the canister and may cause the absorbant
to become water-logged. In top-to-bottom flow absorbers, gravity
leads to more even distribution of condensed water vapor throughout
the absorbant.
Unidirectional
valves
The unidirectional
inspiratory and expiratory valves in most circle absorbers are of
the turret type, in which the pressure generated by the patient's
breathing causes the disc to rise and allow gas to pass in one direction
only. Most have a transparent dome so that the operation of the
valve may be observed.
The disc material
may be mica, ceramic or plastic. Plastic is less expensive, but
tends to warp and allow the valve to become incompetent. Incompetence
may also be caused by the valve sticking in the open position owing
to condensation of water vapor. Incompetent inspiratory or expiratory
valves will reduce the efficiency of gas circulation and result
in rebreathing and consequent carbon dioxide retention.
Some machines
are equipped with valves made of deformable rubber:
As the rubber
ages, these discs tend to harden in a semi-open position, again
allowing the valve to become incompetent.
Connecting
tubing
The body of
the absorber is connected to the patient by means of inspiratory
and expiratory tubes and a Y-piece. This may be constructed of corrugated
black rubber, neoprene or, more recently, plastic.
Recently, the
so-called Universal F circuit has become popular. This is a co-axial
system, the inspiratory tube running inside the expiratory limb:
This arrangement
aids warming and humidification of the inspired gases, albeit at
the expense of an increase in inspiratory resistance to breathing.
One problem with this system, as with other co-axial circuits, is
that, if the inner tube breaks or becomes disconnected at the absorber
end (which may not be noticed on casual inspection), the entire
volume of the tube becomes apparatus dead space. It should also
be noted that, in all other aspects, this system is identical in
function to a conventional, dual-limb system, and does not provide
an economical alternative to the Bain system (although it is occasionally
marketed as doing so).
Patient Size
Most circle
absorbers constructed for use in human-sized animals are satisfactory
for use in patients weighing up to around 100 kg. For larger animals
(e.g. horses and cattle), purpose-built large animal absorbers should
be used. These have larger absorber canisters and larger tubing
and valves in order to reduce resistance to breathing.
The major problem with using standard circle
absorbers in very small animals is that of dead space. First, animals
with very small tidal volumes may not generate enough pressure to
open the valves effectively. Secondly,
the effective dead space of the Y-piece is larger than it appears.
Inevitably, some portion of the expired gas is directed down the
inspiratory limb of the circuit, and some portion of the inspired
gas comes from the expiratory limb, and some mixing of inspired
and expired gases occurs. This may be avoided by placing the valves
in the Y-piece itself, but this arrangement is clumsy and little
used.
|
|
|
(a)
Apparent dead space
|
(b)
Expired gas
|
(c)
Effective dead space
|
These difficulties
may be reduced by the use of purpose-built infant absorbers, which
are smaller than the standard models. Pediatric tubing and Y-pieces,
which are simply smaller in diameter than the standard type, may
be helpful. Purpose-built fans may also be used to circulate the
gases, but these are not comonly encountered in veterinary practice.
In- and out-of-circuit
vaporizers
An inhalation
anesthetic agent may be supplied from a vaporizer positioned in
the circle itself (vaporizer in circuit, VIC) or in the fresh gas
flow from the anesthetic machine (vaporizer out of circuit, VOC).
- Vaporizer
in circuit arrangements have a low-resistance
vaporizer placed in the inspiratory limb of the circle, from
which the anesthetic agent is vaporized by the gases circulated
around the system by the animal's breathing.
VIC
systems are still used in veterinary practice since they employ
an inexpensive vaporizer and provide some degree of autoregulation
of the anesthetic concentration. If the plane of anesthesia becomes
too light, respiration will be less depressed, minute volume will
increase, more agent will be vaporized and the plane of anesthesia
will deepen. It is, however, found that this is not very reliable
in practice.
Although low-resistance vaporizers are usually
relatively inefficient (with the output of, for example, halothane
limited to around 2.5 to 3%), the concentration of anesthetic inspired
by the patient may be very much higher than this because the gas
entering the vaporizer also contains anesthetic from previous
circulations. Indeed, after full equilibration of the circuit and
patient, the inspired concentration would equal the SVP of the anesthetic,
although obviously the patient would have expired long before this
point was reached. It is therefore strongly recommended that an
inhalation anesthetic analyzer be used to monitor the inspired concentration
whenever such systems are used.
In-circuit vaporizers can be used with closed
or semi-closed systems.
Since water vapor exhaled by the patient condenses
in the vaporizer, it is necessary to drain in-circuit vaporizers
regularly.
- Vaporizer
out of circuit systems have the considerable advantage that
a precision vaporizer
may be used to introduce a precisely-known concentration of anesthetic
into the circuit. However, since this delivered concentration
is diluted by the gas already contained in the circuit, the concentration
inspired by the patient is not known with certainty.
The
rate of change of anesthetic concentration in the circuit depends
upon the fresh gas flow rate: a high fresh gas flow rate will
achieve equilibration much faster than if a low fresh gas flow
rate is used:
The above
graph shows anesthetic concentrations in a typical small-animal
circle system as a fraction of the out-of-circuit vaporizer concentration,
using different fresh gas flow rates (0.2 to 4 l/min). When low
flow rates are employed, the concentration of anesthetic in the
circuit will change very slowly, which may cause difficulty in
maintaining a satisfactory plane of anesthesia.
Out of circuit vaporizers are usually used
in semi-closed systems: the low fresh gas flow rate required in
closed systems usually makes their use impractical.
Double-canister
absorbers
Many absorbers
designed for use in human patients employ two canisters placed in
series: the top canister is exposed to the expired gases first and
removes most of the carbon dioxide. Any remaining carbon dioxide
is then removed by the bottom canister. When the top canister is
exhausted, the absorbent is discarded, the bottom canister is placed
in the top position and a canister with fresh absorbent is inserted
underneath it. This
arrangement provides optimal efficiency and economy in carbon dioxide
absorption. However, these absorbers are bulkier, heavier and more
expensive than single-canister models.
If this
type of absorber is used, it is a false economy to fill only one
of the two canisters. The soda-lime will be exhausted at the same
rate, efficiency of absorption will be reduced, and the greater
volume of the circuit will delay equilibration of the gas in the
circuit with the fresh gas supplied from the anesthetic machine.
This will not only slow induction and recovery, but will also tend
to increase consumption of the inhalation anesthetic.
Operational
requirements
- The volume
of the breathing bag must be greater than the patient's inspiratory
capacity. This is usually estimated at 30 ml/kg body weight.
- Since soda
lime contains 50% - 70% air around the granules, the volume of
the absorber canister should be at least double that of the tidal
volume of the patient for optimal efficiency.
- Fresh
gas flow requirements--see separate article for closed
and semi-closed systems.
- The normal
circle system is not suitable for very small animals (less than
about 10 kg) since their tidal and minute volumes are too small
to circulate gas effectively.
Advantages
of the circle system
- Economy
of anesthetic consumption.
- Warming
and humidification of the inspired gases.
- Reduced
atmospheric pollution.
- More efficient
use of soda lime than in Waters' canister.
Disadvantages
- Expensive
and rather bulky.
- Unstable
if used closed.
- Slow changes
in the inspired anesthetic concentration with low flows and out-of-circuit
vaporizer.
- The soda
lime and valves in the system increase resistance to breathing.
- Inhalation
of soda-lime dust.
Uses
- Small animals
over 10 kg, large animals.
Waters'
canister
|